MW640-15 MWIR Lenses

https://www.optotherm.com/web/image/product.template/4534/image_1920?unique=60aecad

Lenses for the MW640-15 MWIR camera for microscopic temperature measurement and analysis.
Product Information: https://www.optotherm.com/shop/4534


Explore this product

Overview

 We developed four radiometric MW640-15 lenses, enabling microscopic temperature measurement of features down to 2µm.  Lenses are manufactured from Germanium, Zinc Selenide, and Zinc Sulfide with anti-reflective coatings for maximum transmission in the long wave infrared region.  Each lens is designed for maximum sensitivity, long working distance, and with minimum distortion for clear, precise imaging.  The threaded interface reduces the space required to attach the lens while maximizing thermal uniformity between the lens and camera body to maintain the high measurement accuracy.

Example Images

Click on the images below to view the actual 640 x 512p image. Reference pixel resolution of each lens image using the USAF table below.

Applications

  • Microscopic temperature measurement and analysis of features down to 2µm in size

Features

Comprehensive Calibration

Each lens is calibrated while installed on the camera, resulting in unsurpassed accuracy and measurement uniformity across the entire field-of-view.

Specifications

Optics

      

MW640-15 Relay Lens

MW640-15 80 Micron Lens

MW640-15 20 Micron Lens

MW640-15 5 Micron Lens

MW640-15 2 Micron Lens

Optical Magnification

1X

0.1875X

0.75X

3X

7.5X

Pixel Resolution

N.A.

80 µm

20 µm

5 µm

2 µm

Focus

N.A.

Fixed

Fixed

Fixed

Fixed

Working Distance

10 mm

195 mm

68mm

27 mm

13 mm

Field of View

N.A.

51.2 x 40.96 mm

12.8 x 10.24 mm

3.2 x 2.56 mm

1.28 x 1.024 mm

Focal Length

23.8 mm

35.7 mm

40.1 mm

34.6 mm

19.8 mm

Object Space NA

0.126

0.025

0.098

0.367

0.600

Depth of Field

±50µm

±2000µm

±100µm

±10µm

±2µm


Measurement

Calibration Range †

N.A.

Low range: 10 to 80°C *

High range: 10 to 300°C

Low range: 10 to 80°C *

High range: 10 to 300°C

Low range: 10 to 80°C *

High range: 10 to 300°C

Low range: 10 to 80°C *

High range: 10 to 300°C

Sensitivity (NETD) ‡

N.A.

Low range: < 20 mK (100 Hz)

Low range: < 20 mK (100 Hz)

Low range: < 20 mK (100 Hz)

Low range: < 20 mK (100 Hz)

Accuracy

N.A.

Low range: +/-2°C

High range: +/-2°C or 2% of reading §

Low range: +/-2°C

High range: +/-2°C or 2% of reading §

Low range: +/-2°C

High range: +/-2°C or 2% of reading §

Low range: +/-2°C

High range: +/-2°C or 2% of reading §

* Sentris low range calibrations are optimized for highest sensitivity and may have reduced calibration range.

† The calibration range stated in the specification indicates the range of black-body temperatures used during calibration. Temperature measurements outside of the stated range may be able to be performed but may not meet the specified measurement accuracy.

‡ Noise Equivalent Temperature Difference (NETD) specifies the smallest temperature difference that can be detected. The value stated is NETD at 100 Hz frame rate. Activating real-time image averaging can improve NETD but will increase the measurement response time.

§ whichever is greater